Huber LCS 80

SKU:
VDH32100163
  • Description

    Huber LCS 80

    Chiller with air-cooled refrigerating unit and circulation pump (stainless steel). Housing, atmospheric open tank and copper soldered evaporator made of stainless steel. With digital level indicator. Condenser for outdoor operation in air-cooled design, performance-optimized by a built-in high-efficiency fan motor. Powerful feed pump with integrated overtemperature protection. The flow rate can be adjusted via the manual bypass valve on inside of the chiller. Control unit B400 / RB400: Remote Control unit capacitive operating interface with OLED display and multi-coloured status notification for instant identification of the current operating status. Choice of eight different system languages (DE, EN, ES, FR, IT, PT, RU, TR). Separate operating option for the feed pump and the cooling unit with convenient adjustment of the desired setpoint. Operating of the system can be evaluated on a PC or notebook via an integrated RS232 interface.

    Technical data according to DIN 12876 Operating temperature range -45...+15 °C Temperature stability at -40°C ±2.0 K temperature set point / display colour LED Touchscreen Internal temperature sensor Pt100 Interface digital RS232 Safety classification Class I / NFL Remote Controller 20m length of remote cable Ambient temperature +35°C +40°C Cooling power at+15°C 20 kW at -15°C 20 kW at -20°C 18 kW at -25°C 16 kW at -30°C 12 kW at -40°C 8,5 kW 8 kW Refrigeration machine air-cooled, CFC- and HCFCfree Refrigerant R449A Refrigerant quantity 16 kg Gas warning sensor without Pressure pump max. delivery 53 l/min max. delivery pressure 4,7 bar Pump connection G 1” male Bath volume min. 40 l Bath volume fill level advance warning 75 l Bath volume max. 105 l sound pressure level +/- 4 dB(A) 77 dB(A) Overall dimensions WxDxH ** 2015x1100x2000 mm Order-No.: VDH32100163 Net weight 880 kg Power supply requirement 460V 3~ 60Hz max. current 50 A Fuse on power network side 3x 60 A Degree of Protection IP54 max. ambient temperature +50 °C min. ambient temperature -20 °C

    Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original. Accessories and periphery: 2pcs Hose nozzles ∅25mm*, bath cover*, Bypass valve*, remote control*, drain valve, thermofluid, RS232 cable * standard accessories Output data valid for: Room temperature 20°C to 50°C. If the ambient temperature rises, the cooling capacity may drop. Special for Outdoor:  100% of rated cooling power up to +35°C ambient temperature  Reduction of cooling power above +35°C ambient temperature  Suitable for outside mounting and winter operation (min. ambient temperature -20°C)  Delivered with remote control with 20m cable  Protection classification of the electrical components IP54  Max. ambient temperature +50°C In accordance with EN60034-1 the following voltage and frequency tolerances are valid: Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example: 5% voltage and +2% frequency -> not allowed! - 5% voltage and - 2% frequency -> allowed Information to Electromagnetic compatibility: Classification (disturbance) to EN55011: Class A, Group 1 Recommended thermofluid: Temper -55 (Potassium Formate Heat Transfer Fluid) Standard delivery conditions - Power cable configuration: 1. Single-phase devices (115V) -> with cable and plug 2. Two-phase devices (208V) -> with cable, without plug 3. Three-phase devices (208V/460V) with current consumption less than 63A -> with cable, without plug 4. Three-phase devices (208V/460V) with current consumption greater than 63A -> with cable, without plug