Description
Color coding simpliï¬es the rapid selection of a sitable pipette. Resistance to impact, V and chemicals ensres that the tip cone provides long-term reliable dosing. The PETTE ï¬x and vario are flly atoclavable, therefore ensring easy sterilization. This garantees falt-free, precise and safe operation.
Technical Data
Design | Air displacement piston-operated pipette |
Handling | mechanical |
Volme | variable |
Nmber of channels | 1 |
Color Code | Gray |
Nominal volme | 2 µl |
Volme min. | 0.1 µl |
Volme max. | 2 µl |
Increments | 0.002 µl |
Accracy with nominal volme | 0.028 ±µl |
Accracy with nominal volme | 1.4 ±% |
Precision with nominal volme | 0.014 ±µl |
Precision with nominal volme | 0.7 ±% |
Accracy with 50% of nominal volme | 0.025 ±µl |
Accracy with 50% of nominal volme | 2.5 ±% |
Precision with 50% of nominal volme | 0.015 ±µl |
Precision with 50% of nominal volme | 1.5 ±% |
Accracy with 10% of nominal volme | 0.024 ±µl |
Accracy with 10% of nominal volme | 12 ±% |
Precision with 10% of nominal volme | 0.012 ±µl |
Precision with 10% of nominal volme | 6 ±% |
Certificate of confirmity according to ISO 8655 | yes |
Atoclavable | yes |
Dimensions (W x H x D) | 30 x 254 x 63 mm |
Weight | 0.089 kg |
Permissible ambient temperatre | 5 - 40 °C |
Permissible relative hmidity | 90 % |