PLEASE NOTE: We will be closed Monday 11.11.2024 in observance of Veterans' Day.

Humidity Chamber 56L HCP 50 115V

SKU:
HCP 50 115V
  • Description

    Humidity Chamber HCP50

     

    Active humidity control creates a controlled environment for accelerated life tests and 85/85 tests.

    Temperature

    Working-temperature range +7 above ambient temperature up to +90 °C
    Setting temperature range +18 to +90 °C
    Setting accuracy temperature 0.1 °C
    Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

    Standard equipment

    Works calibration certificate Works calibration certificate (measuring point chamber centre) at 60 °C with 75 % rh
    Entry port Entry port (silicone), 40 mm clear diameter, for introducing connections, moisture tight, can be closed by silicone stopper, at the back, centre left
    Door Additional heated inner glass door
    Door fully insulated stainless steel door with 2-point locking (compression door lock)
    Internals 1 perforated stainless steel shelf/shelves

    Humidity

    Humidity Control Active microprocessor control for humidifying and dehumidifying (20 – 95 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7; from an external tank) by a self-priming pump; integral bacteria block by generating hot steam, dehumidifying via sterile filter
    Setting range active humidity control 20 to 95 % rh and rh-Off
    Setting accuracy humidity 0.5 % rh

    Stainless steel interior

    Dimensions w(A) x h(B) x d(C): 400 x 425 x 330 mm (d less 35 mm for fan)
    Interior W.-St. 1.4301 (ASTM 304)
    Volume 56 l
    Max. number of internals 5
    Max. loading of chamber 75 kg
    Max. loading per internal 15 kg

    Control technology

    ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
    Language setting German, English, Spanish, French, Polish, Czech, Hungarian
    adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime
    Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
    Function SetpointWAIT the process time does not start until the set temperature is reached
    Calibration three freely selectable values each, temperature and humidity

    Textured stainless steel casing

    Dimensions w(D) x h(E) x d(F): 559 x 795 x 521 mm (d +56mm door handle)

    Electrical data

    Voltage Electrical load 230 V, 50/60 Hz
    approx. 1520 W
    Voltage Electrical load 115 V, 50/60 Hz
    approx. 1520 W

    Ambient conditions

    Ambient temperature 10 °C to 30 °C
    Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
    Humidity rh max. 70 %, non-condensing
    Altitude of installation max. 2,000 m above sea level
    Overvoltage category II
    Pollution degree 2

    Communication

    Interface Ethernet LAN, USB
    Documentation programme stored in case of power failure
    Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

    Safety

    AutoSAFETY additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
    Autodiagnostic system integral fault diagnostics for temperature and humidity control
    Alarm visual and acoustic

    Heating concept

    6 sides large-area multi-function heating system on four sides with additional door and back heating to avoid condensation

    Packing/shipping data

    Transport information The appliances must be transported upright
    Customs tariff number 8419 8998
    Country of origin Federal Republic of Germany
    WEEE-Reg.-No. DE 66812464
    Dimensions approx incl. carton w x h x d: 730 x 950 x 640 mm
    Net weight approx. 55 kg