Huber Minichiller 600-H OLÉ

SKU:
3066.0003.98
  • $9,750.00
  • Description

    Huber Minichiller 600-H OLÉ

    The compact Minichillers are an economic and environmentally friendly cooling solution for many laboratory applications. A calculation of water and waste water costs shows that a Minichiller can save up to 48,000 litres of water in a working week and due to its low procurement price, return on investment is achieved within a very short time. The units have small footprints, only requiring 225 x 360 mm on the laboratory bench and are therefore suitable for cooling reactor blocks, vacuum pumps, rotary evaporators or heat exchangers. Minichillers allow stable application conditions due to constant values for temperature, pressure and throughflow quantities. They are also perfect for continuous operation at ambient temperatures up to 40°C. Despite the economic purchase price the coolers are well equipped and have a large temperature indicator, fill level indicator, as well as status LEDs for the pump, cooling and heating. A 1 kW heater is optionally available with which the temperature range can be extended to +100°C. Minichillers are available as both air and water cooled models and use natural refrigerants.
     The new OLÉ controller combines state-of-the-art technology with simple operation. OLÉ models are suitable for routine tasks in research and industry and are convincing as practice oriented basic equipment. Option: Pt100 sensor connection #10519 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).
    Environmentally-friendly, economic cooling solution
    High efficiency
    Precise temperature control
    Simple operation with menu navigation
    Robust stainless construction
    Reliable continuously operation with warning functions
    Easy to fill and empty
    Illuminated filling level indicator
    Bright, large OLED display
    USB and RS232 interfaces
    Natural refrigerant

     
    General information & temperature controller 
    Temperature range  -20...100 °C
    Temperature stability ±0.2 °C
    Temperature display  OLED Display
    Display resolution  0.1 °C
    Alarm message  optical, acoustical
    Controller OLÉ 
    Safety class  III / FL
    Protection class  IP20
    Dimensions (W x D x H) 280 x 490 x 424 mm | 11.02 x 19.29 x 16.69 inches
    Weight 36 kg | 79.38 pounds
    Heating / cooling capacity 
    Heating capacity  1,6 - 2,1 kW
    Cooling capacity
    15 0 -10 -20  °C
    0,6 0,5 0,35 0,15  kW
    Cooling machine air-cooled
    Refrigerant (ASHRAE, GHS) R290 (A3, H220) (0,085 kg)
    Circulation pump  
    Pressure pump max.  24 lpm ; 0.7 bar | 6.34 gpm ; 10.15 psi
    Suction pump max.  18 lpm ; 0.4 bar | 4.76 gpm ; 5.8 psi
    Pump connection  M16x1 AG male
    Connections 
    Digital interface  USB (Device), RS232 Interface
    Operating data
    Permissible ambient temp. 5...40 °C
    Noise level 56 dB(A)
    Filling volume min.  2.8 Liters | 0.74 gal
    Volume of expansion 2.2 Liters | 0.58 gal
    Power supply 208-240V 1~/2~ 50/60Hz
    Current max. 12 A
    Fusing 16 A
    Miscellaneous
    Please take further technical information as well as information about the scope of delivery, accessories and peripheral devices from the PDF data sheet. In cases of doubt the details given in there apply. If nothing to the contrary is quoted then the performance data referred is at an ambient temperature of 20°C.

    Please observe the notes on installation and/or the space requirements of our products.

    The following voltage and frequency tolerances apply based on EN60034-1:
    Voltage +/- 5% for a frequency tolerance at the same time of +/- 2%

    Examples to illustrate this: Not permissible, for example. is -5% on the voltage and +2% on the frequency. Permissible, on the other hand, are tolerances of -5% on the voltage and -2% on the frequency.

    This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration. Certification by a Notified Body upon request.

    We reserve the right to make technical changes and correct errors. Illustrations can deviate from the original.