Forced Air Circulation (Fan) Incubator, 256L, IF 260 Plus 115V

SKU:
IF 260 PLUS 115V
  • $7,615.00
  • Description

    IF260plus

    Standard delivery Incubator

    The heating of this incubator is optimally tuned for forced air circulation; the fan can also be switched off completely, and valuable chamber loads for research, pharmaceutics, medicine and food chemistry are warmed up very carefully.

    Temperature

    Setting temperature range +20 to +80 °C
    Working temperature range min. 10°C above ambient up to +80°C
    Setting accuracy temperature 0.1 °C
    Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

    Control technology

    ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
    Language setting German, English, Spanish, French, Polish, Czech, Hungarian
    Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
    Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
    Function SetpointWAIT the process time does not start until the set temperature is reached
    Calibration three freely selectable temperature values
    adjustable parameters temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
    Sterilisation fixed sterilisation programme (4 hours/160°C) for sterilisation of working chamber, not for sterilising the load

    Ventilation

    Fan forced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually
    Fresh air Admixture of pre-heated fresh air by electronically adjustable air flap
    Vent vent connection with restrictor flap

    Communication

    Documentation programme stored in case of power failure
    Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

    Safety

    Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
    Temperature control overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
    AutoSAFETY additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
    Autodiagnostic system for fault analysis
    Alarm visual and acoustic

    Standard equipment

    Works calibration certificate incl. works calibration certificate for +37°C
    Door fully insulated stainless steel door with 2-point locking (compression door lock)
    Door inner glass door
    Internals 2 stainless steel grid(s), electropolished

    Stainless steel interior

    Dimensions w(A) x h(B) x d(C): 640 x 800 x 500 mm (d less 39 mm for fan)
    Interior easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
    Volume 256 l
    Max. number of internals 9
    Max. loading of chamber 300 kg
    Max. loading per internal 20 kg

    Ambient conditions

    Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
    Altitude of installation max. 2,000 m above sea level
    Ambient temperature +5 °C to +40 °C
    Humidity rh max. 80 %, non-condensing
    Overvoltage category II
    Pollution degree 2