Rotronic
AwTherm Body incl. Standard AwT-MHS and AwT-PS14 and AwT-PS40, PS14, PS40, HW4-P-QUICK, EA00, EA10, EA35, EA80, EA95, USB Cable-Type A/B
- SKU:
- AwTherm-SET
Description
Temperature-Stabilized Water Activity Analyzer - Rotronic AW-Therm
Reliable Measurement of AW in Food, Cosmetics or Pharma Industries
AwTherm Set Includes:
- AwTherm Body- AwT-MHS
- Sample Holder AwT-PS14
- Sample Holder AwT-PS40
- Sample Cups PS14
- Sample Cups PS40
- HW4-P-QUICK Software
- Humidity Standards: EA00, EA10, EA35, EA80, EA95
- USB Cable-Type A/B
Features
- Accuracy: ±0.005 aw, ±0.1 K
- Temperature control range: 0...60 °C
- High temperature stability: ±0.01 °C/min
- Variable sample container sizes: 14 / 40 mm
- Interchangeable reference probe for cleaning and calibration
- AW Quick function for fast measurement results (typically 4 – 5 minutes)
- Real-time clock with backup battery
- Touch panel for simple operation
The AwTherm from Rotronic is a high-end laboratory instrument for water activity measurements in the food, cosmetics and pharmaceutical industries. The significant advantage of the AwTherm lies in the possibility to heat or cool the measurement chamber, which simplifies and optimizes the measurement process considerably in the case of tempered products, be they goods in cold storage or on heated conveyor belts.
Accuracy | ±0,005 aw (10...30 °C) ±0,1 °C (±0,18 °F) |
---|---|
Chamber-temperature gradient | <0,1 °C (<0,18 °F) |
Operating range | 1...40 °C (34...107 °F) |
Sample container dimensions | AwT-PS14: Ø46 x 14 mm AwT-PS40: Ø46 x 40 mm |
Sample sizes | Variable (14 mm / 40 mm) |
Device type | Temperature-stabilized water-activity measurement |
Measuring range | 0.005...1.000 aw |
Temperature-control range | 0...60 °C (32...140 °F) |
Temperature stability | ±0,01 °C/min (±0,018 °F/min) |
Power consumption | 110...230 V / 50...60 Hz ≤2 A |
Long-term stability | <0.01 aw/year |
Humidity sensor | Hygromer IN-1 |
Temperature sensor | PT100 1/3 DIN Class B |
Calibration / adjustment | Aw & Temperature: HG2-S (probe removed with AwT-CAL) Aw: ROTRONIC humidity standards (via HygroSoft) |
Display type | 8-line LCD with touch operation |
HygroSoft compatible | Yes |
Trend indicator | Yes |
Aw-quick mode | Yes |
Software Update | Via USB port |
Communication interfaces | Micro USB |
Housing | PC / ABS |
IP protection rating | IP21 |
Standards | ISO 18787 |
CE / EMC conform | EMC 2004/108/EC IEC EN 61010-1:2010 |
Maintenance | Annual calibration (recommended) |
Dimensions | 400 x 180 x 180 mm |
Weight | 4200 g |