Description
Ashing Furnace LV5/11, up to 1100C, 240V
Ashing furnace LV 5/11 is designed especially for ashing processes to 1050 °C in the laboratory. Applications include determining loss on ignition, ashing food and plastics for subsequent substance analysis. A special fresh-air and exhaust air system ensures that the air is replaced 6 times per minute so that there is always sufficient oxygen for the ashing process. Incoming air passes the furnace heating and is pre-heated to ensure good temperature uniformity.
Standard Equipment
- Tmax 1100 °C
- Heating from two sides
- Ceramic heating plates with integral heating element which is safeguarded, and easy to replace
- Air exchange of more than 6 times per minute
- Good temperature uniformity due to preheating of incoming air, temperature uniformity according to DIN 17052-1 to +/− 10 °C in the defined empty work area (from 550 °C)
- Suitable for many standardized ashing processes according to ISO, ASTM, EN, and DIN
- Optional flap door (LV) which can be used as work platform or lift door (LVT) with hot surface facing away from the operator
- Controller with touch operation B510 (5 programs with each 4 segments)
Model Flap door |
Tmax in °C1 |
Inner dimensions in mm |
Volume in l |
Outer dimensions2 in mm |
Max. weight of hydrocarbons in g |
Max. evapo- ration rate g/min |
Max. connected load in kW |
Electrical connection* |
Weight in kg |
Heating time in min4 |
||||
w |
d |
h |
W |
D |
H3 |
|||||||||
LV 3/11 |
1100 |
180 |
150 |
120 |
3 |
345 |
390 |
810 |
5 |
0.1 |
1.3 |
1-phase |
20 |
45 |
LV 5/11 |
1100 |
205 |
170 |
130 |
5 |
385 |
415 |
810 |
10 |
0.2 |
2.6 |
1-phase |
29 |
55 |
LV 9/11 |
1100 |
235 |
240 |
170 |
9 |
415 |
485 |
865 |
15 |
0.3 |
3.3 |
1-phase |
36 |
70 |
LV 15/11 |
1100 |
230 |
340 |
170 |
15 |
415 |
590 |
865 |
25 |
0.3 |
3.6 |
1-phase |
44 |
80 |